Robocount 70+
Features
- Cost effective 2 Regents system with low Regent consumption
- Internal preheating of diluent for every test.
- Mouse & keypad connection by USB.
- Supports LIS & HIS
- Clog detection & prevention with automatic system cleaning process.
- External Icey PAD/mouse by USB, Supports USB data backup
- Four kinds of QC modes: L-J, X, X-R and X-B
- Build-in thermal printer & option for external printer
SPECIFICATIONS
PARAMETERS
- WBC, LYM#, MID#, GRAN#, LYM%, MID%,
- GRAN%, RBC, HGB, HCT, MCW, MCH, MCHC, RDW-CV,
- RDW-SD, PLT, MPV, PDV-SD, PDW-CV, PCT, P-LCR, P-LCC
- 3-part differentiation of WBC, 22 parameters and 3 histograms
PRINCIPLES OF OPERATION
- WBC/RBC/PLT: electrical impedance
- HGB: photoelectric colorimetry
MEASUREMENT
- 2 counting chambers
- Aperture diameter WBC 100 μm RBC/PLT : 68 μm
- 72 Samples/hour Throughput
Regents & sampling system
- Sample: 13 ul for whole blood, 20 uL for Pre-diluted mode
- Regents: 2 Regent’s: Diluent & Lyse
Technical features
- 10.4 Inch color LCD display
- Data storage upto 100000 sample results with histograms
- Built-in thermal printer
- External Icey pad & mouse
- Support Rs 232, 6 USB (mouse, keyboard, HP printer, barcode reader (optional) & Hard disk)
Control & calibration
- L-J, X-R, X-B four control modes, manual and auto-calibration
Operating temperatures
- 15-35°C, optional temperature 25°C
Power requirement
- AC 100-240V, 50/60 Hz, 180 VD
Dimensions